Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 863: CAL ADJ /5700,5725 INSTRUMENT: Fluke 865: CAL ADJ /5700,5725 INSTRUMENT: Fluke 867: CAL ADJ /5700,5725 DATE: 21-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 348 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # # Source: # Fluke 86X Calibration Procedure, Preliminary, February 1995 # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # 1.001 MMFC C2 1.002 ASK- W 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Plug external power adapter into UUT to insure stable 1.005 DISP power while calibration is being performed. 1.006 DISP If at any time, you wish to abort the calibration, 1.006 DISP select the MET/CAL Terminate option and "Abort CAL" 1.006 DISP (Soft Key 5) on the UUT. 1.007 DISP 10 minute warm-up time requirement. 1.007 DISP Ambient temperature: 18C - 28C. 1.007 DISP Relative humidity: less than 90%. 1.008 HEAD STEP 1: mVDC CALIBRATION ADJUSTMENT 1.009 DISP Connect the 5700A to the UUT as follows: 1.009 DISP [32] 5700A OUTPUT HI to UUT Volt Ohm Diode Input 1.009 DISP [32] 5700A OUTPUT LO to UUT COM Input 1.009 DISP [32] 5700A AUX CURRENT OUTPUT to UUT mA/uA Input 1.010 DISP Rotate the UUT function switch to mVDC. 1.010 DISP 1.010 DISP Place the UUT in calibration mode by pressing the. 1.010 DISP recessed calibration button located below the power 1.010 DISP adapter plug on the right side of the UUT. 1.010 DISP 1.010 DISP Select Advance when the calibration display appears. 1.011 5700 900mV S 2W 1.012 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.013 DISP Select Advance when Step 2 appears on the left side of 1.013 DISP the UUT display. 1.014 HEAD STEP 2: mVDC CALIBRATION ADJUSTMENT 1.015 5700 2.9V S 2W 1.016 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.017 DISP Select Advance when Step 3 appears on the left side of 1.017 DISP the UUT display. 1.018 HEAD STEP 3: mVDC CALIBRATION ADJUSTMENT 1.019 5700 290mV S 2W 1.020 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.021 DISP Select Advance when Step 4 appears on the left side of 1.021 DISP the UUT display. 1.022 5700 * S 1.023 HEAD STEP 4: VDC CALIBRATION ADJUSTMENT 1.024 DISP Rotate the UUT function switch to VDC. 1.024 DISP 1.024 DISP Select Advance when the calibration display appears. 1.025 5700 1000V S 2W 1.026 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.027 DISP Select Advance when Step 5 appears on the left side of 1.027 DISP the UUT display. 1.028 HEAD STEP 5: VDC CALIBRATION ADJUSTMENT 1.029 5700 290V S 2W 1.030 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.031 DISP Select Advance when Step 6 appears on the left side of 1.031 DISP the UUT display. 1.032 HEAD STEP 6: VDC CALIBRATION ADJUSTMENT 1.033 5700 29V S 2W 1.034 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.035 DISP Select Advance when Step 7 appears on the left side of 1.035 DISP the UUT display. 1.036 5700 * S 1.037 HEAD STEP 7: mADC CALIBRATION ADJUSTMENT 1.038 DISP Rotate the UUT function switch to mA/uA. 1.038 DISP 1.038 DISP Select Advance when the calibration display appears. 1.039 5700 290mA S 2W 1.040 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.041 DISP Select Advance when Step 8 appears on the left side of 1.041 DISP the UUT display. 1.042 HEAD STEP 8: mADC CALIBRATION ADJUSTMENT 1.043 5700 29mA S 2W 1.044 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.045 DISP Select Advance when Step 9 appears on the left side of 1.045 DISP the UUT display. 1.046 HEAD STEP 9: mADC CALIBRATION ADJUSTMENT 1.047 5700 2.9mA S 2W 1.048 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.049 DISP Select Advance when Step 10 appears on the left side of 1.049 DISP the UUT display. 1.050 HEAD STEP 10: mADC CALIBRATION ADJUSTMENT 1.051 5700 290uA S 2W 1.052 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.053 DISP Select Advance when Step 11 appears on the left side of 1.053 DISP the UUT display. 1.054 HEAD STEP 11: mADC CALIBRATION ADJUSTMENT 1.055 5700 29uA S 2W 1.056 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.057 DISP Select Advance when Step 12 appears on the left side of 1.057 DISP the UUT display. 1.058 5700 * S 1.059 DISP Disconnect the 5700A from the UUT. 1.060 HEAD STEP 12: ADC CALIBRATION ADJUSTMENT 1.061 DISP Connect the 5725A to the UUT as follows: 1.061 DISP [32] 5725A CURRENT OUTPUT HI to A Input 1.061 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM Input 1.062 DISP Rotate the UUT function switch to A. 1.062 DISP 1.062 DISP Select Advance when the calibration display appears. 1.063 5700 10A B1 S 2W 1.064 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.065 DISP Select Advance when Step 13 appears on the left side of 1.065 DISP the UUT display. 1.066 HEAD STEP 13: ADC CALIBRATION ADJUSTMENT 1.067 5700 2.9A B1 S 2W 1.068 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.069 DISP Select Advance when Step 14 appears on the left side of 1.069 DISP the UUT display. 1.070 HEAD STEP 14: ADC CALIBRATION ADJUSTMENT 1.071 5700 0.29A B1 S 2W 1.072 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.073 DISP Select Advance when Step 15 appears on the left side of 1.073 DISP the UUT display. 1.074 5700 * S 1.075 DISP Disconnect the 5725A from the UUT. 1.076 HEAD STEP 15: Ohms/nS CALIBRATION ADJUSTMENT 1.077 DISP Connect the 5700A to the UUT as follows: 1.077 DISP [32] 5700A SENSE HI to UUT Volt Ohm Diode Input 1.077 DISP [32] 5700A SENSE LO to UUT COM Input 1.077 DISP [32] 5700A OUTPUT HI to UUT Volt Ohm Diode Input 1.077 DISP [32] 5700A OUTPUT LO to UUT COM Input 1.077 DISP [32] 5700A AUX CURRENT OUTPUT to UUT mA/uA Input 1.078 DISP Rotate the UUT function switch to Ohms. 1.078 DISP 1.078 DISP Select Advance when the calibration display appears. 1.079 5700 1.9MZ S 2W 1.080 DISP Press Modify Value (Soft Key 2). 1.081 DISP Use the Next Digit (Soft Key 1) and the up and down 1.081 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.081 DISP [MEM1] MOhms on the UUT display. 1.081 DISP 1.081 DISP Select Done (Soft Key 5) when ready to proceed. 1.082 DISP Press Proceed (Soft Key 1). 1.083 DISP Select Advance when Step 16 appears on the left side of 1.083 DISP the UUT display. 1.084 HEAD STEP 16: Ohms/nS CALIBRATION ADJUSTMENT 1.085 5700 190kZ S 2W 1.086 DISP Press Modify Value (Soft Key 2). 1.087 DISP Use the Next Digit (Soft Key 1) and the up and down 1.087 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.087 DISP [MEM1] kOhms on the UUT display. 1.087 DISP 1.087 DISP Select Done (Soft Key 5) when ready to proceed. 1.088 DISP Press Proceed (Soft Key 1). 1.089 DISP Select Advance when Step 17 appears on the left side of 1.089 DISP the UUT display. 1.090 HEAD STEP 17: Ohms/nS CALIBRATION ADJUSTMENT 1.091 5700 19kZ S CW 1.092 DISP Press Modify Value (Soft Key 2). 1.093 DISP Use the Next Digit (Soft Key 1) and the up and down 1.093 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.093 DISP [MEM1] kOhms on the UUT display. 1.093 DISP 1.093 DISP Select Done (Soft Key 5) when ready to proceed. 1.094 DISP Press Proceed (Soft Key 1). 1.095 DISP Select Advance when Step 18 appears on the left side of 1.095 DISP the UUT display. 1.096 HEAD STEP 18: Ohms/nS CALIBRATION ADJUSTMENT 1.097 5700 1.9kZ S CW 1.098 DISP Press Modify Value (Soft Key 2). 1.099 DISP Use the Next Digit (Soft Key 1) and the up and down 1.099 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.099 DISP [MEM1] kOhms on the UUT display. 1.099 DISP 1.099 DISP Select Done (Soft Key 5) when ready to proceed. 1.100 DISP Press Proceed (Soft Key 1). 1.101 DISP Select Advance when Step 19 appears on the left side of 1.101 DISP the UUT display. 1.102 HEAD STEP 19: Ohms/nS CALIBRATION ADJUSTMENT 1.103 5700 190Z S CW 1.104 DISP Press Modify Value (Soft Key 2). 1.105 DISP Use the Next Digit (Soft Key 1) and the up and down 1.105 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.105 DISP [MEM1] Ohms on the UUT display. 1.105 DISP 1.105 DISP Select Done (Soft Key 5) when ready to proceed. 1.106 DISP Press Proceed (Soft Key 1). 1.107 DISP Select Advance when Step 20 appears on the left side of 1.107 DISP the UUT display. 1.108 HEAD STEP 20: Ohms/nS CALIBRATION ADJUSTMENT 1.109 5700 100nY S 2W 1.110 DISP Press Modify Value (Soft Key 2). 1.111 DISP Use the Next Digit (Soft Key 1) and the up and down 1.111 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.111 DISP [MEM1]nS on the UUT display. 1.111 DISP 1.111 DISP Select Done (Soft Key 5) when ready to proceed. 1.112 DISP Press Proceed (Soft Key 1). 1.113 DISP Select Advance when Step 21 appears on the left side of 1.113 DISP the UUT display. 1.114 HEAD STEP 21: Ohms/nS CALIBRATION ADJUSTMENT 1.115 5700 1000nY S 2W 1.116 DISP Press Modify Value (Soft Key 2). 1.117 DISP Use the Next Digit (Soft Key 1) and the up and down 1.117 DISP arrows (Soft Keys 2 and 3) to enter a value of 1.117 DISP [MEM1]nS on the UUT display. 1.117 DISP 1.117 DISP Select Done (Soft Key 5) when ready to proceed. 1.118 DISP Press Proceed (Soft Key 1). 1.119 DISP Select Advance when Step 22 appears on the left side of 1.119 DISP the UUT display. 1.120 5700 * S 1.121 HEAD STEP 22: VAC CALIBRATION ADJUSTMENT 1.122 DISP Rotate the UUT function switch to VAC. 1.122 DISP 1.122 DISP Select Advance when the calibration display appears. 1.123 5700 1000V 1kH S 2W 1.124 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.125 DISP Select Advance when Step 23 appears on the left side of 1.125 DISP the UUT display. 1.126 HEAD STEP 23: VAC CALIBRATION ADJUSTMENT 1.127 5700 100V 1kH S 2W 1.128 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.129 DISP Select Advance when Step 24 appears on the left side of 1.129 DISP the UUT display. 1.130 HEAD STEP 24: VAC CALIBRATION ADJUSTMENT 1.131 5700 290V 1kH S 2W 1.132 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.133 DISP Select Advance when Step 25 appears on the left side of 1.133 DISP the UUT display. 1.134 HEAD STEP 25: VAC CALIBRATION ADJUSTMENT 1.135 5700 29V 1kH S 2W 1.136 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.137 DISP Select Advance when Step 26 appears on the left side of 1.137 DISP the UUT display. 1.138 HEAD STEP 26: VAC CALIBRATION ADJUSTMENT 1.139 5700 2.9V 1kH S 2W 1.140 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.141 DISP Select Advance when Step 27 appears on the left side of 1.141 DISP the UUT display. 1.142 HEAD STEP 27: VAC CALIBRATION ADJUSTMENT 1.143 5700 290mV 1kH S 2W 1.144 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.145 DISP Select Advance when Step 28 appears on the left side of 1.145 DISP the UUT display. 1.146 HEAD STEP 28: VAC CALIBRATION ADJUSTMENT 1.147 5700 29mV 1kH S 2W 1.148 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.149 DISP Select Advance when Step 29 appears on the left side of 1.149 DISP the UUT display. 1.150 5700 * S 1.151 HEAD STEP 29: mAAC CALIBRATION ADJUSTMENT 1.152 DISP Rotate the UUT function switch to mA/uA. 1.152 DISP 1.152 DISP Select Advance when the calibration display appears. 1.153 5700 290mA 1kH S 2W 1.154 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.155 DISP Select Advance when Step 30 appears on the left side of 1.155 DISP the UUT display. 1.156 HEAD STEP 30: mAAC CALIBRATION ADJUSTMENT 1.157 5700 29mA 1kH S 2W 1.158 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.159 DISP Select Advance when Step 31 appears on the left side of 1.159 DISP the UUT display. 1.160 HEAD STEP 31: mAAC CALIBRATION ADJUSTMENT 1.161 5700 2.9mA 1kH S 2W 1.162 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.163 DISP Select Advance when Step 32 appears on the left side of 1.163 DISP the UUT display. 1.164 HEAD STEP 32: mAAC CALIBRATION ADJUSTMENT 1.165 5700 290uA 1kH S 2W 1.166 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.167 DISP Select Advance when Step 33 appears on the left side of 1.167 DISP the UUT display. 1.168 HEAD STEP 33: mAAC CALIBRATION ADJUSTMENT 1.169 5700 29uA 1kH S 2W 1.170 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.171 DISP Select Advance when Step 34 appears on the left side of 1.171 DISP the UUT display. 1.172 5700 * S 1.173 DISP Disconnect the 5700A from the UUT. 1.174 HEAD STEP 34: AC Amps CALIBRATION ADJUSTMENT 1.175 DISP Connect the 5725A to the UUT as follows: 1.175 DISP [32] 5725A CURRENT OUTPUT HI to A Input 1.175 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM Input 1.176 DISP Rotate the UUT function switch to A. 1.176 DISP 1.176 DISP Select Advance when the calibration display appears. 1.177 5700 10A 1kH B1 S 2W 1.178 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.179 DISP Select Advance when Step 35 appears on the left side of 1.179 DISP the UUT display. 1.180 HEAD STEP 35: AC Amps CALIBRATION ADJUSTMENT 1.181 5700 1A 1kH B1 S 2W 1.182 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.183 DISP Select Advance when Step 36 appears on the left side of 1.183 DISP the UUT display. 1.184 HEAD STEP 36: AC Amps CALIBRATION ADJUSTMENT 1.185 5700 2.9A 1kH B1 S 2W 1.186 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.187 DISP Select Advance when Step 37 appears on the left side of 1.187 DISP the UUT display. 1.188 5700 * S 1.189 DISP Disconnect the 5725A from the UUT. 1.190 HEAD STEP 37: AC Amps CALIBRATION ADJUSTMENT 1.191 DISP Connect the 5700A to the UUT as follows: 1.191 DISP [32] 5700A OUTPUT HI to Volt Ohm Diode Input 1.191 DISP [32] 5700A OUTPUT LO to UUT COM Input 1.191 DISP [32] 5700A AUX CURRENT OUTPUT to UUT A Input 1.192 5700 290mA 1kH S 2W 1.193 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.194 DISP Select Advance when Step 38 appears on the left side of 1.194 DISP the UUT display. 1.195 HEAD STEP 38: mVDC CALIBRATION ADJUSTMENT 1.196 DISP Rotate the UUT function switch to mVDC. 1.196 DISP 1.196 DISP Select Advance when the calibration display appears. 1.197 5700 29mV S 2W 1.198 DISP Press Proceed (Soft Key 1) when the reading has settled. 1.199 DISP Select Save Cal. to store new constants or Abort Cal. 1.199 DISP when calibration completes. 1.200 END